The numerical investigation and experimental results of the proto-pump LAP-00 show that at design rotational speed of 9000 r/min
the flow rate corresponding to the highest efficiency point is 11 L/min
which is much higher than clinical use(38 L/min)
causing the pump to work at off design point in clinic use with a low efficiency.Besides
in the diffuser region
flow separation and recirculation may cause blood stagnation and thrombosis.The improved design for pump LAP-23 decreases the design rotational speed from 9000 r/min to 8000 r/min
thus reduces the high shear stress on blood cells at impeller tip;meanwhile
it decreases the design flow rate to 6 L/min and eliminates flow separation and recirculation in the diffuser region.
Numerical research on effects of blade tip clearances onthe hemocompatibility of artificial heart pump
Study on improvement of the compressor anti-swirl distortion capacity with bowed stator
Orthogonal optimal design of solid fuel formulation for HO/HTPB hybrid rocket motor
Study on matching method for core and bypass mass flux of 2-D convergent nozzle of mixed flow
Effects of throat asymmetry on micro nozzle's performance
Related Author
ZHANG Yan
LIU Yangwei
TANG Yumeng
XIE Nan
HU Jun
LIU Hua
TU Bao-feng
LI Xin-tian
Related Institution
Department of Cardiac Surgery,Fu Wai Hospital,Chinese Academy of Medical Sciences
Collaborative Innovation Center for Advanced Aero-Engine
National Key Laboratory of Science and Technology on Aero-Engine Aero-Thermodynamics,School of Energy and Power Engineering,Beijing University of Aeronautics and Astronautics
Jiangsu Province Key Laboratory of Aerospace Power System, College of Energy and Power Engineering, Nanjing University of Aeronautics and Astronautics
School of Astronautics, Beijing University of Aeronautics and Astronautics