LIU Lin, WANG Weijia, HUO Ju, et al. Portable vascular blocking tension meter based on metal strain gauge[J]. 2021, 53(2): 184-190.
DOI:
LIU Lin, WANG Weijia, HUO Ju, et al. Portable vascular blocking tension meter based on metal strain gauge[J]. 2021, 53(2): 184-190.DOI: 10.11918/202009093.
Portable vascular blocking tension meter based on metal strain gauge
two-vessel occlusion (2-VO) is often used to establish cerebral ischemia models
while due to the non-quantitative ligation of the artery
the model usually has poor reproducibility
large individual differences within the group
and high mortality. In this paper
a vascular blocking tension meter is proposed by taking metal strain gauge as the core component to realize the quantitative detection of ligation force. Biological experimental verification was carried out in vitroand in vivo
as well as in such aspects as cells and molecules level
tissues and organs level
and overall level. Results show that the linearity and sensitivity of the tension meter were affected by factors such as base parameters
bonding process
circuit parameters
and appearance structure of the metal strain gauge. Compared with the non-instrumental ligation group
the animal mortality of the cerebral ischemia model was reduced by 42.22%. The decrease in cerebral blood flow of the experimental animals corresponded to the increase in ligation force. After hematoxylin-eosin (HE) staining of the hippocampus
light microscopy results show that the neurons were loosely arranged
the number of neurons decreased and even disappeared
the cytoplasmic staining of the neurons became weak
and the dentate gyrus exhibited neuronal pyknosis. A comparison between different ligation force groups shows that the individual differences in the hippocampus morphology of the rats in the 1.5 N ligation force group were the smallest. Morris water maze results show that compared with the sham operation group
the escape latencies of the rats in the 1.5 N ligation group and the 2.5 N ligation group were significantly increased (P<0.05)
and the number of platform crossing times in the 1.5 N ligation group was significantly decreased (P<0.05). There was an obvious learning and memory disorder. It was falsified that the greater the ligation force of the 2-VO rats was
the better the ligation effect was
which was obtained in our previous laboratory experiments. The conclusion was drawn that the smaller the standard deviation of the ligation force was
the better the ligation effect was
and the ligation force could be quantitatively controlled. The application of the vascular blocking tension meter to the basic experimental research of medicine is of significance.