DUAN Yun, WANG Rang-ding, ZHU Ying, et al. A Method for Reducing the Static Drift of the Time Difference Method Ultrasonic Flowmeter[J]. 2010, 27(8): 205-209.
DUAN Yun, WANG Rang-ding, ZHU Ying, et al. A Method for Reducing the Static Drift of the Time Difference Method Ultrasonic Flowmeter[J]. 2010, 27(8): 205-209.DOI:
The large static drift will deteriorate dynamic measurement accuracy and precision of the time difference method ultrasonic flowmeter
while accurate determination of the received ultrasonic signal can efficiently reduce the drift of the static instantaneous flow. The traditional hardware threshold model is relatively weak in anti-interference
and there is large static drift
which seriously affects the measurement precision improvement. In this paper
the zero threshold smoothing model is improved a lot on the threshold model so that it can accurately determinate the received ultrasonic signal. The measured results prove that this method can efficiently reduce the drift of the static instantaneous flow
and improve the precision of the time difference method ultrasonic flowmeter.