Based on analysis and improvement of simple response method(SRM) and Weibull reliability method for calibration intervals of test system
and by taking into account fault decisions on the calibration process
the decision risks on verification of test system were resolved.In this paper
related national measurement standards on reliability index of test system were firstly introduced
then some usual calibration interval determination methods were also discussed
and furthermore the reliability probability functions for observational value of test reliability were given.By synthesizing different parameters with principal component analysis(PCA) method
the estimation values of the optimal calibration interval were corrected.Lastly
the results of verification on digital multimeter(DMM) measurement channel in automatic test system(ATS) show that the calibration intervals optimization methods are feasible.